EasyManua.ls Logo

BEAMEX MC5-IS - Page 115

BEAMEX MC5-IS
164 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
107
3. When the calibration run is
ready, you can either Reject
the results (A/Reject Calibra-
tion) or Continue by pressing
the D/Saveè Function Key.
Note.
If the maximum error of the
calibration results exceed the
Adjust if > error limit, a mes-
sage window with the follow-
ing text:
The calibrated instrument
need to be adjusted.
4. Enter the environmental data
and save the results using one
of the options provided in the
Function Keys B to D.
Note.
Some of the temperature data
may be pre-entered by MC5-
IS, provided the data is avail-
able for MC5-IS (pressure
modules, the E module, the
ET module).
5. Continue either with another
calibration run or end the cali-
bration procedure. In the lat-
ter case you will see the pos-
sible Finishing Guide text before you are back viewing the
Instrument Data.
Note.
If you are calibrating instruments using off-line method then please
keep in mind this: When you upload the results from MC5-IS to the
calibration software, the results are by default also left in MC5-IS’
memory. To free memory in MC5-IS by deleting the results, see
Deleting Instruments on page 123.
See also:
MC5-IS’ Support for Instrument Adjustment on page 118.
Viewing Calibration Results on page 124.
25.09.2000 10:52
Input
Error
1.00
Output
Pt100 α385
°C (ITS90)
4.0293
0.18 % of span
0.00
+
0
-
0% 50% 100%
RTD Temperature [ET: Simul]
Current [E: Meas]
mA
Calibration
Reject
Save
25.09.2000 10:55
ENVIRONMENTAL DATA
Device Temperature
Environment Temp.
Input Module Temp.
Output Module Temp.
Humidity
Calibrated by:
Notes:
Andy Handyman
Just Testing
21.0
21.0
31.6
29.6
°C
°C
°C
°C
30
%
As Found
Save
As Left
Save
As Both
Save
Back
Calibrating an Instrument

Table of Contents

Related product manuals