153
Safety Information
Connections Using E Module Terminals Only
Simple connections,
i.e. between COM (14) and one other terminal.
Combined connections (applies to all other
connections in the E module except the ones
mentioned in the table above).
Other Type of Front Panel Connections
For all other type of front panel connections: Add the parameters of all simple connections that
are part of the intended combined connection.
Ex Approvals
Special Conditions for Safe Use
The permissible ambient temperature range is -20 °C ≤ T
a
≥ + 50 °C.
The calibrator shall only be used with rechargeable battery pack type
RB8412-IS.
The battery pack shall not be charged in a hazardous location.
Use only Beamex charger type BC14-IS for charging.
The RS232 connection can be used in safe area only.
The EXT-IS connection shall only be connected with the EXT-IS module,
external pressure calibration module.
The calibrator may be used with auxiliary hand held mechanical pressure
calibrator pump type PGM.
The surface of the external pressure module and the calibrator pump con-
tains light metal, so the user shall be aware of possible impact or friction
sparks when using these auxiliary devices.
The connection cable of the external pressure module as well as the pres-
sure measurement hose are non-conducting material, so the user shall be
aware of dangers of ignition due to hazardous electrostatic charges.
Due to electrostatic discharge hazard, do not affix any stickers larger than
4 cm
2
(0.62 in
2
) to the calibrator.