60
Startup and Basic Operation
Voltage Generation
The ET module has terminals for voltage generation within -2.5 …+10 V range and termi-
nals for low voltage sensor simulation with the range ±250 mV. The low voltage terminals
are also used when measuring/simulating thermocouples using an external Reference
Junction.
Required settings Options/description
Quantity Voltage
Funct/Port ET: V(gen) or ET: LowV(sim)
Generating Voltages between +10 V
and -2.5 V
Select Function/port ET: V(gen) and
choose a suitable unit.
MC5-IS displays both the set value and the
internally measured value for the generated
voltage in the selected window. Max. out-
put current is 1 mA.
Low Voltage Generation
Select Function/port ET: LowV(sim) and
choose a suitable unit. The max. output
current is 5 mA.
MC5-IS displays both the set value and the
internally measured value for the generated
voltage in the selected window.
T / C , L o w V
H A R T
®
1
m e a s / s i n k
V , ,
V ,
1
,
s i n k
C o m
R , R T D
L o w V
M E A S U R EO U T P U T
E T E
T / C I N T . R J
S E N S O R M E A S U R E & S I M U L A T E
U o : 1 6 V
I o : 2 4 m A
P o : 1 4 5 m W
T / C
W I R E S
O N L Y
4 w m e a s
3 w m e a s
m e a s / s i m
T / C , L o w V
T / C , L o w V
H A R T
®
1
m e a s / s i n k
V , ,
V ,
1
,
s i n k
C o m
R , R T D
L o w V
M E A S U R EO U T P U T
E T E
T / C I N T . R J
S E N S O R M E A S U R E & S I M U L A T E
U o : 1 6 V
I o : 2 4 m A
P o : 1 4 5 m W
T / C
W I R E S
O N L Y
4 w m e a s
3 w m e a s
m e a s / s i m
s i n k
V ,
1
,