EasyManua.ls Logo

BEAMEX MC5-IS - Viewing Calibration Results

BEAMEX MC5-IS
164 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
116
Calibration
Electrical Limit Switches
This procedure suits limit switches with an electrical input. The in-
put signal is generated with MC5-IS.
Required modules
The ET module for generating the required electrical signal
(Voltage or Current).
The E module for detecting the switch’s state.
Preparations
1. Connect the ET module’s terminals
marked OUTPUT” to the input of the
switch. Add a current source to the loop.
MC5-IS controls the current provided by
the current source.
2. Connect the switch contact to the switch
terminals in the E module.
3. Test the connections in Basic Mode if
needed. To quickly configure the Basic
Mode’s windows, go to Calibration
Mode, select the instrument to be cali-
brated and the Function Key C/Cali-
brate but immediately return to Basic
Mode.
Calibration
1. Move to Calibration Mode and select the
instrument to be calibrated.
2. Start the calibration as is presented in
chapter A Calibration Procedure Us-
ing MC5-IS on page 106. MC5-IS does
the Prescan if it is enabled. During the
prescan, MC5-IS searches for approxi-
mate values for the actuating and
deactuating point. This speeds up the
final test without sacrificing accuracy.
The prescan test is done only once in a
calibration.
By default, the Prescan is set to on.
If you do not want MC5-IS to per-
form a prescan, disable it (com-
mands D/MENU and 2/Prescan).
When Prescan is set to off, MC5-
IS uses the whole scan range also
during the actual test.
3. The actual test is done automatically:
MC5-IS slowly increases the input sig-
nal until the switch actuates and con-
tinues by decreasing the input signal
until the switch deactuates. MC5-IS’
screen displays the obtained data as
the test advances.
06.10.2000 10:29
Input
Output
V
8.0310
5.023
Voltage [ET: Gen.]
Switch [E: Meas.]
Actuate
Pause
Testing switch limits (4/5)
Deactuate-------
7.000
5.500
4.500
3.000
Current
Source
MULTIFUNCTION CALIBRATOR
T/C, Low V
HART
®
I
meas/sink
V, ,
V, I ,
sink
Com
R, RTD
?
Low V
MEASUREOUTPUT
ET E
T/C INT. RJ
SENSOR MEASURE & SIMULATE
Uo: 16 V
Io: 24 mA
Po: 145 mW
T/C
WIRES
ONLY
II 1 G
EEx ia IIC T4
(Tamb = 50°C)
VTT 02 ATEX 003 X
Ex ia IIC T4
(Tamb = 50°C)
VTT No. Ex-02.011 X
Ui: 30 V
Ii: 215 mA
Pi: 1 W
INPUT
4 w meas
3 w measmeas/sim

Table of Contents

Related product manuals