58
Startup and Basic Operation
Current Sink
MC5-IS has a possibility to sink current, i.e. MC5-IS acts as a device controlling the flow
of current while a external device generates the supply voltage for the loop, both in the E
and the ET module. The maximum current it 25 mA.
Required settings Options/description
Quantity Current
Funct/Port ET: I(control) or E: I(control)
Using the ET module’s output termi-
nals
When sinking current using the ET
module’s output terminals, select Funct/
Port option ET: I(control). The following
picture displays the active terminals:
Note.
If you are simultaneously using the ET module’s output terminals
and some of the ET module’s measuring terminals, there is no gal-
vanic isolation between the measuring circuit and the current sink-
ing circuit.
Using the E module’s output termi-
nals
Select Funct/Port option E: I(control) to
sink current using the E modules terminals.
Active terminals:
Note.
The output current of the E module is a “pure” analog signal without
any embedded HART signal.
T / C , L o w V
H A R T
®
1
m e a s / s i n k
V , ,
V ,
1
,
s i n k
C o m
R , R T D
L o w V
M E A S U R EO U T P U T
E T E
T / C I N T . R J
S E N S O R M E A S U R E & S I M U L A T E
U o : 1 6 V
I o : 2 4 m A
P o : 1 4 5 m W
T / C
W I R E S
O N L Y
4 w m e a s
3 w m e a s
m e a s / s i m
s i n k
V ,
1
,
T / C , L o w V
H A R T
®
1
m e a s / s i n k
V , ,
V ,
1
,
s i n k
C o m
R , R T D
L o w V
M E A S U R EO U T P U T
E T E
T / C I N T . R J
S E N S O R M E A S U R E & S I M U L A T E
U o : 1 6 V
I o : 2 4 m A
P o : 1 4 5 m W
T / C
W I R E S
O N L Y
4 w m e a s
3 w m e a s
m e a s / s i m
1
m e a s / s i n k
C o m