40
Startup and Basic Operation
Voltage Measurement
The E module has terminals for low voltage measurement within the range ±250 mV and
terminals for voltage measurement within ±30 V range. The ET module also has low volt-
age measurement terminals with a range of ±250 mV. The ET module terminals are also
used when measuring/simulating thermocouples using an external Reference Junction.
Required settings Options/description
Quantity Voltage
Funct/Port ET: LowV(mea), E: LowV(mea) or E: V(meas)
MC5-IS displays the measured low voltage in the selected window.
Hint!
Low voltage measurement can be used for non-standard thermo-
couple measurement. You will see the measured temperature in
millivolts and need a table to convert the measured millivolt value to
corresponding temperature values. In this case, use copper exten-
sion cords to connect the non-standard thermocouple to MC5-IS’
terminals.
Measuring Low Voltages
Select either Function/port ET: LowV(mea)
or E: LowV(mea) and choose a suitable
unit. The following pictures display the ac-
tive terminals for both available ports.
ET: LowV(mea), ±250 mV:
E: LowV(mea), ±250 mV:
T / C , L o w V
H A R T
®
1
m e a s / s i n k
V , ,
V ,
1
,
s i n k
C o m
R , R T D
L o w V
M E A S U R EO U T P U T
E T E
T / C I N T . R J
S E N S O R M E A S U R E & S I M U L A T E
U o : 1 6 V
I o : 2 4 m A
P o : 1 4 5 m W
T / C
W I R E S
O N L Y
4 w m e a s
3 w m e a s
m e a s / s i m
T / C , L o w V
T / C , L o w V
H A R T
®
1
m e a s / s i n k
V , ,
V ,
1
,
s i n k
C o m
R , R T D
L o w V
M E A S U R EO U T P U T
E T E
T / C I N T . R J
S E N S O R M E A S U R E & S I M U L A T E
U o : 1 6 V
I o : 2 4 m A
P o : 1 4 5 m W
T / C
W I R E S
O N L Y
4 w m e a s
3 w m e a s
m e a s / s i m
C o m
L o w V