EasyManua.ls Logo

BEAMEX MC5-IS - Scaling

BEAMEX MC5-IS
164 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
78
Advanced Operation and Configuration
Scaling
In scaling, the measuremed/generated/simulated value is dis-
played in another, user defined quantity/unit. Start scaling as de-
scribed in chapter Display Mode Settings on page 77.
When selecting Scaling Display Mode, the following configuration
window appears:
SCALING
26.04.2002 14:02
Quantity
Port
Pressure
P1:INT2C
Measured Range
kPa
Stop
Ramping
Field StartCancel Edit
0 %
100 %
kPa
0.000
100.000
Scaled Mode
Scaled Unit
mA
Scaled Range
0 %
100 %
4.000
20.000
Transfer Function Linear
Current
SCALING
30.05.2002 10:09
Quantity
Port
Pressure
P1:INT2C
Measured Range
kPa
Stop
Ramping
Field StartCancel Edit
0 %
100 %
kPa
0.000
100.000
Scaled Mode
Custom Units
Scaled Unit
Scaled Range
0 %
100 %
1.00
5.00
Transfer Function Linear
Gallons/min
Decimals
3
You can scale the quantity either to a quantity that already exists in
MC5-IS or select Custom Units
from the Scale Mode field. Then
you can manually enter any unit to
the Scale Unit field (see the
leftmost picture above).
When scaling is active in a Basic
Mode window, the window con-
tains the text “SCALED”. The true
value is displayed in Extra Info row.
If the Extra Info row was assigned
to some other measurement be-
fore scaling was started, the other
measurement is automatically
stopped.
Note.
Remember to set the amount of decimals to be displayed for a
Custom Unit. The default value is 3.
30.05.2002 10:10
P2: INT20C/-1.0 ... 20.7 bar g
E: Current Measurement
Gallons/min
20.088
8.3217
mA
2
Current
1
Pressure
Measurement bar
1.004
Mode
Calibration
MENU
SCALED

Table of Contents

Related product manuals