Loading...
10
General
Batteries
MC5-IS uses rechargeable batteries that are specially designed for
use in this intrinsically safe calibrator.
The charger for rechargeable batteries operates in the following
environments:
- Voltage: 100 … 240 VAC,
- Frequency: 50/60 Hz
Warning!
The charger may be used in safe area only!
The charging electronics is in the Battery Pack. Therefore the bat-
teries may be charged although the Battery Pack is disconnected
from MC5-IS’ Base Unit. If you have two sets of rechargeable bat-
teries you may charge the disconnected Battery Pack while at the
same time use MC5-IS with the connected Battery Pack.
The maximum operating time without recharging varies depending
on the usage MC5-IS. A good average operating time is 5 hours.
The upper left corner of MC5-IS’ display shows a picture of a bat-
tery. The whiter the picture is, the more acute is the need for re-
charging.
Notes.
MC5-IS’ memory and the internal clock/calendar uses a small
amount of power although the calibrator is switched off. Remember
to check the capacity of the batteries from time to time although
MC5-IS is not in use.
Do not leave MC5-IS without a Battery Pack for a long time. MC5-IS
may lose its settings if it is left without a support voltage for an
extended period.
Warning!
Use only the batteries designed for use in MC5-IS. Use of other
batteries may cause unpredictable hazards resulting in dam-
ages, injuries and even death.
Do not charge the batteries in hazardous area.
Full batteries:
Empty batteries: