111
Calibration
1. Move to Calibration Mode and select the
instrument to be calibrated.
2. Start the calibration as presented in
chapter A Calibration Procedure Us-
ing MC5-IS on page 106.
3. Set the pressure manually using, e.g.
a pressure pump.
4. Accept the points using the B/Accept
Readings Function Key.
5. Save or reject the results as is de-
scribed in chapter A Calibration Pro-
cedure Using MC5-IS on page 106.
6. Either do another calibration run or re-
turn to the Instrument window.
Hint!
These instructions can be adapted for pneumatic pressure trans-
mitters and converters, too. Instead of connecting the output sig-
nal to a terminal in the E module, connect the pressure output sig-
nal to a suitable pressure module.
Warning!
Working with pressure instruments can be hazardous. Only
qualified personnel may use pressure instruments and pres-
sure sources.
Next…
MC5-IS’ Support for Instrument Adjustment on page 118
Calibrating an Instrument