Installation Operating instructions CS 300
4 - 14 6.110.98.5.01.40 en
4
Mounting the weighing system CS 300 MA 84
The scanner must not be connected.
–
Unplug scanner.
– Please observe the scanner operating instructions .
CAUTION No correct cable routing when inserting the weighing system into
a scanner!
Cables get into contact with mobile parts of the weighing system.
A shunt force is generated and the weighing results are not cor-
rect.
–
Route the cables in a way that no mobile parts get into contact with
the system.
Connect scanner cable to adapter.
Fig. 24: Scanner cable
Adapter
Scanner cable
ZIF plug
Connect the scanner cable
to the adapter
with the ZIF plug
. Open fuse of
ZIF plug, insert cable and close fuse.