700M HFID 6-2-15
161
AKEN: Device Identification
Command
Response
Description
_AKEN_K0
_AKEN_K1
_AKEN_K2
_AKEN_K3
_AKEN_K4
_AKEN_K5
_AKEN_s_devicename
_AKEN_s_model
_AKEN_s_serial no
_AKEN_s_Air pressure
_AKEN_s_Sample pressure
_AKEN_s_Fuel pressure
Device identification is responded.
Device Model
Device Serial Number
Air Pressure
Sample Pressure
Fuel pressure
ATEM: Temperatures
Command
Response
Description
_ATEM_K0
_ATEM_s_z.z_y.y_x.x……
1. Filter Temperature
2. Burner Temperature
3. Oven Temperature
4. Cutter Temperature
5. Pump Temperature
_ATEM_K0_x
_ATEM_S_xtemp
Temperature of x
ADRU: Pressures / Electronic Pressure Control Valve Voltage in Percent
Command
Response
Description
_ADRU_K0
_ADRU_s_z.z_y.y_x.x……
1. Sample Pressure
2. Air Pressure
3. Fuel Pressure
4. Air Inject Pressure
5. Fuel Inject Pressure
6. % of Sample EPC Volts
7. % of Air EPC Volts
8. % of Fuel EPC Volts
9. % of Air Inject EPC Volts
10. % of Fuel Inject EPC Volts
_ADRU_K0_x
_ADRU_S_xpress/x% voltage
Reading of x