Site Selection and Mounting
CAUTION: The following precautions must be carefully observed:
1. Select a site free from direct sunlight, radiation from a high-temperature surface, or
abrupt temperature variations.
2. This analyzer is not suitable for installation outdoors.
3. Select a site where the air is clean. Avoid exposing the instrument to corrosive or
combustible gases.
4. Do not subject the analyzer to severe vibration. If severe vibration is present, use
isolation mounts.
5. The instrument is designed for rack mounting. Optional rack-mount slides are
available.
6. Do not install the 700M HFID near equipment that emits electromagnetic interference
(EMI).
NOTE: A front and rear supporting brace or equivalent are required if the optional rack mount
slides were not purchased.
The Power On/Off switch is accessible from the rear of the analyzer only. DO NOT
mount the analyzer in a manner that leaves the Power On/Off switch inaccessible.