Liberty
™
Automated Microwave Peptide Synthesizer
600178 • Revision 5 • October 2009
68
A warning box will pop up about the reaction vessel. Verify that the reaction 3.
vessel is properly connected to the Liberty, and then click OK to continue.
A warning box will pop up about the contents of the reaction vessel. Verify that 4.
there are no solid reagents in the reaction vessel, and then click OK to continue.
The Liberty will perform an unltered drain on the reaction vessel to ensure it is 5.
empty, and then it will purge all liquid from all manifolds. This process will take
approximately 5 minutes.
The System Check window will open. As each test is completed, the values 6.
will be recorded in the window. The System Check will take approximately 20
minutes to complete.
The results of the7. System Check will be rendered as a PDF. Click Save to save
the PDF.