Importing and Exporting Sensor Calibration Values
To create a backup copy of the current sensor calibration values:
From the Sensor Calibration tab, click Export.1.
Enter a name for the backup le, and then click Save.2.
To restore previous sensor calibration values from a backup le:
From the Sensor Calibration tab, click Import.1.
Select the backup le to be restored, and then click Open.2.
Pressure Calibration Tab
The Pressure Calibration tab allows for the calibration of the pressure sensor on the Liberty.
Pressure calibration is not necessary for routine maintenance; a calibration should only be done
as part of a troubleshooting protocol as advised by CEM Service.
NOTE
When importing values from a backup le, the current calibration values for
all sensors will be discarded.