Liberty
™
Automated Microwave Peptide Synthesizer
600178 • Revision 5 • October 2009
79
Importing and Exporting Sensor Calibration Values
To create a backup copy of the current sensor calibration values:
From the Sensor Calibration tab, click Export.1.
Enter a name for the backup le, and then click Save.2.
To restore previous sensor calibration values from a backup le:
From the Sensor Calibration tab, click Import.1.
Select the backup le to be restored, and then click Open.2.
Pressure Calibration Tab
The Pressure Calibration tab allows for the calibration of the pressure sensor on the Liberty.
Pressure calibration is not necessary for routine maintenance; a calibration should only be done
as part of a troubleshooting protocol as advised by CEM Service.
NOTE
When importing values from a backup le, the current calibration values for
all sensors will be discarded.