EasyManua.ls Logo

CEM liberty - Performing a Large Scale Calibration

CEM liberty
118 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Liberty
Automated Microwave Peptide Synthesizer
600178 • Revision 5 • October 2009
73
To restore previous volume calibration values from a backup le:
From the Volume Calibration tab, click Import.1.
Select the backup le to be restored, and then click Open.2.
Performing a Large Scale Calibration
Place the reaction vessel into the 1. volume calibration stand (PN 576250).
Remove the reaction vessel assembly from the cavity of the Discover. 1.1.
Set the volume calibration stand on top of the cavity opening, with the 1.2.
graduations on the right.
Insert the reaction vessel into the calibration stand so that the drain line 1.3.
from the bottom of the vessel ts into the slot on the left of the stand.
From the Volume Calibration Tab, click on the Large Scale Addition Tab.2.
NOTE
When importing values from a backup le, the current calibration values for
all reagents will be discarded.

Table of Contents

Related product manuals