Sample System Check File
Manifold Check
-0.1
Nitrogen:
-0.1
Ext1:
0.0
AA1:
0.0
AA2:
0.0
AA3:
-0.1
Ext2:
0.0
AA4:
Flow Performance
Flow Path
15.9 PSI
Nitrogen:
Reading(PSI)
Delta
Ext1
10.9
4.9
Flow Path
Reading(PSI)
Delta
AA2
12.9
3.0
AA1
12.9
3.0
AA3
12.9
3.0
Ext1ToSprayhead
11.3
4.6
AA4
12.1
3.8
Ext1toFilter
11.6
4.2
Ext1_Ext2ToSprayhead
12.7
3.2
Ext1toExt2
12.7
3.1
Ext2
10.7
5.2
Ext2ToFilter
11.4
4.5
Ext2ToSprayhead
11.1
4.8
FilterTop
10.8
5.1
SprayHead
10.9
5.0
FilterBottom
11.4
4.5
ROT3
11.1
4.8
V2
5.1
10.8
TFAToRV
12.1
3.7
Ext1_2ToFilter
12.8
3.0
0.0
RV:
Ext2Bypass
10.3
5.5
0.0
Ext2 Bypass:
Valve Check
-0.1/-0.1
RV2:
-0.1/-0.1
RV3:
0.0/-0.1
RV4:
0.0/0.0
RV6:
0.0/0.0
RV7:
-0.1/0.0
M2:
0.0
M1:
-0.1/-0.1
M3:
(Normally Closed/Normally Open)
4/24/2009 9:27 AM
4/24/2009 9:52 AM
Start:
End:
Valve RV7(3.1, 3.7) Failure: Valve did not open
CEM Liberty System Check
207A16
Liberty S/N:
DP7094
Firmware:
PepDriver Version:
2.5.4.2
201A11
Discover S/N:
Firmware:
Please email this file to: service@cem.com