Liberty
™
Automated Microwave Peptide Synthesizer
600178 • Revision 5 • October 2009
80
Performing a Pressure Calibration
Click Calibrate to begin the calibration procedure. A warning will pop up about 1.
performing a backpurge. Click Yes to backpurge the system before calibrating.
Adjust the pressure on the main pressure regulator in the Liberty to 16 psi, and 2.
then click OK to continue the calibration.
NOTE
Put empty bottles on all positions before performing a backpurge.