Liberty
™
Automated Microwave Peptide Synthesizer
600178 • Revision 5 • October 2009
81
The Liberty will vent the pressure on the entire system to get a zero pressure 3.
reading, and then apply pressure to get a main pressure reading.
Look at the pressure regulator on the Liberty and enter the current reading. Click 4.
OK to continue the calibration.
The software will save the calibration values automatically. Click Verify to test the 5.
pressure readings.
The Test System screen will automatically open and begin performing a Pressure 6.
Adjustment to conrm the pressure is within the correct range.