Sensor Calibration Tab
The Sensor Calibration tab allows for the calibration of the sensors used to monitor the addition
and transfer of reagents in the Liberty. Sensor calibration should be performed monthly as part of
routine maintenance.
Performing a Sensor Calibration
Open the Sensor Calibration tab. PepDriver will load the current calibration 1.
values for each sensor.
Check the Select All box. This will select all sensors except LS2/3.2.
NOTE
For detailed instructions on performing a sensor calibration, see the Liberty
Maintenance Procedures manual, CEM PN 600119.
NOTE
LS2/3 do not need to be routinely calibrated. A separate calibration
procedure is required for these sensors. Contact CEM Service for help if
these sensors require calibration.