Liberty
™
Automated Microwave Peptide Synthesizer
600178 • Revision 5 • October 2009
77
Click Calibrate to the begin the sensor calibration procedure.3.
A series of warning boxes will pop up to ensure sufcient DMF is loaded on the 4.
Liberty for the test. Verify there is sufcient DMF in the Main Wash bottles and
click OK.
The instrument will pressurize the DMF bottles and then ll all manifolds and 5.
tubing with DMF. Once this is complete, the calibration procedure will begin. As
each sensor is calibrated, the updated calibration value is displayed in the New
Threshold column. The entire process should take approximately 5 minutes.
NOTE
Although the calibration procedure only requires 75 mL of DMF, each bottle
should have a minimum of volume of 150 mL to ensure accurate delivery.