Liberty
Automated Microwave Peptide Synthesizer
600178 • Revision 5 • October 2009
75
Once the addition is nished, measure the volume delivered to the reaction
vessel using the graduations on the calibration stand.
Enter the delivered volume in the Delivered Calibration Volume box, and then 9.
click Calibrate. Two values will appear below the Calibrate button: New Rate and
New Prime Volume. Click Save to save the new calibration values.