EasyManua.ls Logo

Chroma 62000D Series - Setting Time; Setting Voltage; Setting Voltage Slew Rate; Setting Current

Chroma 62000D Series
180 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Program Sequence
4-11
D. Setting sequence type to SKIP
When SEQ TYPE = SKIP is selected, the Sequence page as shown in Figure 4-12
indicates the Sequence will skip automatically and jump to next sequence.
4.1.2.2 Setting Time
1. Tap Figure 4-12 (7) to set the time parameter.
2. The time range for setting is listed in the table below.
Table 4-2
TIME
Min. (Sec)
Max. (Sec)
0.002
15000
4.1.2.3 Setting Voltage
1. Tap Figure 4-12 (1) to set the voltage parameter.
2. See section 3.2 for detail description.
4.1.2.4 Setting Voltage Slew Rate
1. Tap Figure 4-12 (2) to set the voltage slew rate.
2. See section 3.2 for detail description.
4.1.2.5 Setting Current
1. Tap Figure 4-12 (3) (4) to set the current parameter.
2. See section 3.2 for detail description.
4.1.2.6 Setting Current Slew Rate
1. Tap Figure 4-12 (5) to set the current slew rate.
2. See section 3.2 for detail description.
4.1.3 Execution in List Mode
When the sequences are finished for editing, press “ ” to start execution and press
again to abort it.

Table of Contents

Related product manuals