SOUR:DCON:FALL? [{MAX/MIN}]
SOUR:DCON:RISE? It returns the setting.
The output must be OFF for setting.
5.6.2.4 LOAD Subsystem
1. LOAD: CURRent
Sets the Load output current (ampere).
LOAD:CURRent <NRf+>[ suffix]
Refer to individual spec for valid numeric range.
LOAD:CURR 1 It sets the output current to 1 amps.
LOAD:CURR 60.00 It sets the output current to 60.00 amps.
LOAD:CURR? It returns the current.
LOAD:CURR? MAX It returns the max. current can be set.
2. LOAD:CURRent:PROTect:HIGH
Sets the current range for Load over current protection.
LOAD:CURRent:PROTect:HIGH <NRf+>[ suffix]
Refer to individual spec for valid numeric range.
LOAD:CURR:PROT:HIGH 60.0 It sets the high limit to 60A
for current output protection.
LOAD:CURR:PROT:HIGH? [{MAX/MIN}]
LOAD:CURR:PROT:HIGH? It returns the high limit of
current protection.
LOAD:CURR:PROT:HIGH? MAX It returns the max. over
current protection..
Sets the Load power output (watt).
LOAD:POWer <NRf+>[ suffix]
Refer to individual spec for valid numeric range.
LOAD:POW 1 It sets the output power to 1 watt.
LOAD:POW 60.00 It sets the output power to 60 watt.
LOAD:POW? It returns the power
LOAD:POW? MAX It returns the max. power can be
set.
4. LOAD:POWer:PROTect:HIGH
Sets the power range for Load over power protection.
LOAD:POWer:PROTect:HIGH < NRf+>[ suffix]