Manual Operation
3-35
Figure 3-54
Each D-series supply has two DisplayPort female sockets. They need to
be connected no matter if the connection is in series or parallel. There is
no difference between left and right connection ports but connected in
series as shown in Figure 3-53 or Figure 3-54. The communication cable
has a limit of 5 meters for length, do not exceed this limit to avoid
unstable signal. The DisplayPort communication cable connected to the
last DC power supply does not connect back to the first DC power
supply.
If it is necessary to return to single unit mode from series or parallel
mode of operation, the DisplayPort connection must be removed or the
single unit will not work normally.
3.2.4.2.4 Setting Series/Parallel Operation Mode
It is necessary to set SLAVE first and MASTER last when operating
62000D Series supplies in series or parallel mode, or a communication
error may result.
1. In Menu page, tap “Configuration” “Series & Parallel” and the screen will appear as
Figure 3-55.
2. Select Series or Parallel connection mode (keypro is required for Series mode).
3. Select the Type to be Master or Slave.
4. Master+
= this indicates that one Slave unit is connected to the master
5. If Master/Slave must be enable for Series or Parallel operation.