Disinfecting / cleaning / sterilizing
Only products from the list of surface disinfectants may
be used for disinfection. To ensure material compatiblity,
products based on the following agents may be used:
– Aldehydes
– Alcohols
Quaternary ammonium compounds.
The following products are not suitable:
Compounds containing alkylamines
Compounds containing phenols
Halogen-releasing compounds
Strong organic acids
– Oxygen-releasing compounds.
Users in the Federal Republic of Germany are advised to
use disinfectants specified in the current list published by
the German Society for Hygiene and Microbiology
(DGHM). The DGHM list also specifies the active agent
of each disinfectant.
Users in countries in which the DGHM list is not available
are advised to use the agents mentioned above.
Alcohols and agents containing alcohol must not be
allowed to seep into the sample line!
Alcohol can distort the measurement results.
Breathing system
All parts of the breathing system, valve discs, bellows,
bellows container, Y-piece, breathing hoses, water
trap containers, breathing bag, parts of the absorber
and Resutator 2000, parts of the secretion aspirator
and parts of the anaesthetic gas scavenging system
Disinfect in hot liquid disinfectant – in a disinfecting
machine at 93 °C for 10 minutes.
Only use cleaning agent, no alkaline or chlorine-
releasing substances – risk of corrosion!
All parts of the breathing system (except the flow
sensor!), Y-piece, breathing hoses, water trap
containers, breathing bag, parts of the absorber, parts
of the secretion aspirator and parts of the anaesthetic
gas scavenging system
can be sterilized in hot steam at 134 °C.
Surfaces of Julian, compressed gas hoses, cables and
Vapor 19.3 / Vapor 2000 / Devapor
The surface of the apparatus must not be treated with
agents containing alcohol.
Wipe off impurities with a damp disposable cloth.
Disinfect with a wipe-on disinfectant, e.g.
Buraton 10 F
(made by Schülke & Mayr GmbH).
Note the manufacturer's Instructions for Use.
Do not allow any liquids to enter openings in the
apparatus.
Flow-Sensor
Disinfect for approx. 1 hour in 70 % ethanol solution.
Leave the sensor to dry in air for at least 30 minutes,
otherwise the remaining alcohol can damage the
sensor when calibrated.
The flow sensor cannot be autoclaved.
The flow sensor can be reused as long as it can be
calibrated successfully.
The flow sensor must be disposed of as infectious
special waste. It can be incinerated with little pollution
at temperatures of more than 800 °C.