Adapting limits – AutoSet
When the ventilation settings have been made, Julian can
automatically adapt the alarm limits for the minute volume
(MV) and airway pressure (PAW) to the current
ventilation settings in IPPV mode.
Press the softkey »auto set limits«.
The new alarm limits for the set MV and PAW are
calculated by Julian as follows:
To reset individual alarm limits for MV and PAW:
See "Displaying and setting alarm limits"
on page 60.
To reuse all default alarm limits:
See "Configuring in standby mode,
setting default alarm limits" on page 72.
62
Monitoring
Alarms
Upper alarm limit Lower alarm limit
MV
[L/min]
VT x f x 1.4
At least 2.0
VT x f x 0.6
At least 0.5
PAW
[mbar]
(Pmax + 10) PEEP +
Pmax – PEEP
4