Selecting alarms
Turn the rotary knob to select the column »alarms«
and press to confirm.
Default
This function is used to activate the default alarm limits
set in the standby configuration which are effective
whenever the apparatus is switched to standby.
Setting default alarm limits:
Select the »default settings« menu by turning the
rotary knob and press to confirm.
See page 86 for the further procedure.
CO2 alarms on/off
To activate / deactivate all CO2 alarms, including the
apnoea alarm.
Turn the rotary knob to select »CO2 on« or »CO2 off«
and press to confirm.
SpO2 alarms on/off
To activate/deactivate all SpO2 alarms.
Turn the rotary knob to select »SpO2 on« or »SpO2
off« and press to confirm.
Deactivated alarms are indicated by the symbol _ in the
measurement field.
National and European standards require minimum
monitoring with alarm function. These standards may not
be met when the alarm function of the monitoring
parameter etCO2 or SpO2 is deactivated.
72
Monitoring
Configuring in operation
Selecting alarms
O
2
-Sensor
21 Vol.%
gas sensor
IPPV
anesth. agent
calibration
pulse tone
alarm sound
scale
parameters
record
activate monitor defaults
0 1 2 3 4 5 6 7 8 9
1 2 3 4 5 6 7 8 9
adult. child
600 10
1:2
10 0251.0050
O
2
%
L/min
PEEP
mbar
Freq.
1/min
V
T
mL
T
I
:T
E
T
IP
:T
I
%
Pmax
mbar
Freshgas O
2
+ N
2
O
selection
auto man
Standard
on off
SpO
2
on off
HLM on off
Isoflurane
PAW
40
20
0
The anesthetic
agent is automati-
cally selected by
gas identification
settings
alarms
CO
2
"Only for internal use by Draeger. Not applicable as Instructions for Use!"