If an alarm occurs
Alarm messages are assigned to three priority classes
by Julian, depending on their urgency, and identified by
exclamation marks:
Warning !!! = message with top priority
A warning message requires immediate action
Caution !! = message with medium priority
A caution message requires prompt action
Advisory ! = message with lowest priority
An advisory message requires to be noted and action
taken if necessary
If several alarms occur simultaneously, Julian displays the
alarm with the highest priority. The others are not
displayed.
All warning, caution and advisory messages that have
been emitted can be displayed by pressing the
»alarm Info« softkey.
Each time an alarm message is displayed, the alarm lamp
either flashes or lights up continuously and an audible
tone sequence is emitted to indicate the alarm priority
class:
Warning:
1 Red lamp flashes, accompanied by a continuous tone.
Caution:
1 Yellow lamp flashes, accompanied by an intermittent
tone every 30 seconds.
Advisory:
1 Yellow lamp lights up continuously, accompanied by a
single tone.
The alarm tone can be suppressed for two minutes:
2 Press the G key and its yellow LED lights up.
To reactivate the alarm tone:
2 Press the G key and its yellow LED goes out.
A list of alarm messages can be found on page 120,
"Message – Cause – Remedy".