Emergency ventilating bag
Example: Dräger Resutator 2000
Ready for operation.
Central gas supply
Gas pressures:
1 All pressure gauges in the green zone.
Backup gas cylinders
2 Slowly open the cylinder valves.
Check the cylinder pressure gauges:
3 O2 pressure greater than 50 bar,
4 N2O pressure greater than 30 bar.
2 Reclose the cylinder valves.
29
Starting Up
Checking the workstation against the checklist
2
2
4
3
N2O O2
Julian
D
3