Suction system
1 Switch to I.
2 Set suction pressure with rotary knob »Vac.«.
3 Block the secretion viewing window or kink the
suction hose.
4 Measure the suction pressure on the pressure gauge.
1 Switch to 0.
O2 flush
5 Press »O2 +« button.
6 The breathing bag inflates.
7 O2 flows from the Y-piece.
Version with rotary knob for O2 emergency metering
8 Turn on rotary knob »Safety O2« for O2 emergency
metering.
6 The breathing bag inflates.
7 O2 flows from the Y-piece.
8 Turn off rotary knob »Safety O2« for O2 emergency
metering again.