Installing the bellows
● Hold the breathing system from above to prevent it
dropping out and
1 uniformly slide the tightly compressed bellows onto
the connector.
2 Insert the container in the breathing system from
below and turn it clockwise as far as possible.
Filling and installing the absorber
3 Push the insert fully into the absorber.
● Fill the absorber with fresh soda lime
– up to the MAX mark.
Only use Drägersorb 800!
If the breathing system is to remain unused during the
next 24 hours:
● Only fill with soda lime immediately before use!
Soda lime dries out when exposed to ambient air
and sun for a longer period of time. This reduces
the CO2 absorption capacity and the effectiveness
of the indicator.
4 Insert the absorber in the breathing system from
below and turn clockwise as far as possible.
5 Slot in the breathing system – as far as possible.
6 Fold down the writing top.
107
Care
Assembling
MAX
3
4
2
1
5
5
6