87
Configuring in Standby Mode
Setting default values
Adjustment range for default alarm limits
Fixed alarm limits / times:
FiCO2 Fixed upper alarm limit 5 mmHg
(deactivated in MAN/SPONT).
Apnoea pressure after 15 seconds
Apnoea flow after 15 seconds
Apnoea CO2 after 15 seconds
(after 60 seconds in MAN/SPONT)
– – : The default value set on delivery is outside the
monitoring range and the corresponding alarm limit
is therefore switched off.
The newly set default alarm limits are valid whenever
the apparatus is switched on or started up from
standby mode.
Certain alarm limits are switched off automatically in
MAN/SPONT ventilation mode and when using the
external fresh gas outlet:
see tables, alarm limits for MAN/SPONT, page 39
and alarm limits for non-rebreathing systems, page 51.
Alarm limit Adjustment range Default value
set on delivery
SpO
2
[%]
51 to 100
50 to 99
– –
92
Pulse
H
[1/min]
31 to 300
30 to 299
120
50
etCO2
[mmHg]
1 to 75
0 to 74
50 (mmHg)
– –
MV
[L/min]
0.1 to 39
0 to 38.9
12
3.0
FiO2
[Vol.%]
19 to 100
18 to 99
– –
20
Fi Hal.
[Vol.%]
0.1 to 7
0 to 6.9
1.5
– –
Fi Iso.
[Vol.%]
0.1 to 7
0 to 6.9
2.3
– –
Fi Enf.
[Vol.%]
0.1 to 7
0 to 6.9
3.4
– –
Fi Des.
[Vol.%]
0.1 to 21.9
0 to 21.8
12.0
– –
Fi Sev.
[Vol.%]
0.1 to 9.9
0 to 9.8
3.4
– –
PAW
[mbar]
5 to 98
0 to 35
40
8
_
_
_
_
_
_
_
_
_
_
_
"Only for internal use by Draeger. Not applicable as Instructions for Use!"