Record
This function is used to determine the type of event that
triggers an entry in the record list or a printout on the
connected logging printer.
● Select the line »record« with the rotary knob and
press to confirm.
The »record« menu appears on the left-hand side.
Display (example):
Time
interval-started (min)
A list entry is activated after a fixed time interval in
minutes.
NiBP-started
An entry is made after every NiBP measurement with new
measured values. This entry can be read by an external
Dräger monitor via the MEDIBUS interface.
warning-started
An entry is made whenever an alarm is triggered.
caution-started
An entry is made whenever a caution message is
triggered.
● Turn the rotary knob to select the event triggering the
entry and press to confirm.
Activating monitor defaults
This function is used to activate the default monitoring
settings configured in standby mode and active whenever
the apparatus is switched on.
● Select the line »activate monitor defaults« with the
rotary knob and press to confirm.
IPPV
0 1 2 3 4 5 6 7 8 9
1 2 3 4 5 6 7 8 9
adult child
600 10
1:2
10 0251.0050
O
2
%
L/min
PEEP
mbar
Freq.
1/min
V
T
mL
T
I
:T
E
T
IP
:T
I
%
Pmax
mbar
Freshgas O
2
+ N
2
O
PAW
40
20
0
settings
puls tone
alarm sound
Scale
parameter
record
activate monitor defaults
Menu for configuration of measurement parameters.
list entry:
time
intervall (min)
NiBP started
warning started
caution started
printer entry:
time
intervall (min)
NiBP started
warning started
caution started
12510 off
on off
on off
on
off
12510 off
on
off
on off
on
off
anesth. agent
calibration
alarms
71
Monitoring
Configuring in operation
Setting the monitoring functions