Select the appropriate sensor.
On the back of the Julian:
1 Swing the flap over the socket »SpO2« to the left and
insert the sensor plug.
C-lock-ECG synchronization (optional)
If the patient is highly agitated or the arterial flow is very
low, the measurement signals of the SpO2 measurement
can be improved with the C-Lock-ECG synchronization.
In this case, Julian receives two separate signals that
transmit the heart activity:
An optical signal from the SpO2 sensor
and
An electrical signal from the separate ECG monitor.
Julian uses the R-wave of the ECG signal to identify the
pulse and to synchronize with the SpO2 measurement.
2 Channel the ECG signal of the ECG monitor with the
cable andyesck-plug to the back of the Julian
Sync.« socket).
See the "Technical data" on page 131 for the
requirements for the electrical signal and plug layout.
In the event of a delayed ECG signal
Synchronization may be impaired if the ECG signal is
delayed by more than 40 milliseconds to the QRS
complex.
If you suspect a fault of this kind:
Use the Julian without C-Lock-ECG synchronization.