Care list for Julian anaesthetic workstation
Only applies for non-infectious patients.
When used with infectious patients, all parts in contact with breathing gas must additionally be
sterilized after disinfection and cleaning.
104
Care
Care list
What How often
Conditioning intervals
1)
Filter on
Y-piece
Filter on inspiration
and expiration port
Without filter
Julian workstation
Vapor 19.3 / Devapor unit Daily
Power cable, compressed gas hoses,
equipotential bonding lead
Monthly
Breathing bag with twin connector
and hose
Daily Daily Per patient
Breathing hoses Daily Per patient Per patient
Bellows Weekly Weekly Daily
Y-piece Daily Per patient Per patient
Cover of breathing system with
APL valve
Weekly Weekly
Daily
Middle and bottom part of breathing
system
Weekly Weekly Daily
Container of bellows Weekly Weekly Weekly
Valve discs (use a cassette) Weekly Weekly Daily
Expiration port Weekly Weekly Daily
Absorber and insert Weekly Weekly Daily
Water trap (monitoring) Daily Daily Daily
Microbial filter 654 St ––– Daily –––
Filter of the measured gas
recirculation system
Weekly Weekly Weekly
Flow sensor Weekly Weekly Daily
AGS housing
Weekly
AGS flow tube (without filter) Weekly
Container for AGS buffer volume Weekly
AGS transfer hose Weekly
Scavenging hose with connector Monthly
Waste gas connector with tube Weekly
Hose of sample gas return or hose of
sample gas scavenging
Weekly
Emergency breathing bag, its
diaphragm, upper and lower valve
parts (Dräger Resutator 2000)
After each use
Silicone sleeve of secretion collecting
bottle and rinsing bottle, their lid with
float, suction hose and inspection
window
As required,
but at least daily
1)
The conditioning intervals depend on the use and position of the filters. The table is
merely intended as a rough guide. The instructions of the hospital's hygiene officer shall
prevail!
Components which
can be conditioned
Front daily, other surfaces weekly