The parts should preferably be cleaned and disinfected in a washing machine, otherwise by
immersion.
Julian and its components must not be treated with formaldehyde vapours or ethylene oxide!
Care
Care list
Sterilization
Washing machine
2)
93 °C 10 minutes
Wiping
3)
Immersion
3)
Steam, 134 °C,
10 minutes
no outside no no
no outside no no
no
yes no no
yes no possible yes
yes no possible yes
yes no possible yes
yes no possible yes
yes no possible yes
yes no possible yes
yes no possible yes
yes no possible yes
yes no possible yes
yes no possible yes
yes no possible yes
no outside
4)
no yes, max. 24 times
no outside no yes, max. 10 times
no outside possible
5)
no
yes no possible no
no yes no no
yes no possible yes
yes no possible no
yes no possible no
yes no possible no
yes no possible yes
yes no possible yes
yes no possible yes
2)
Only use cleaning
agents, no alkaline or
chlorine-releasing
substances. Risk of
corrosion!
3)
Surface disinfectants
based on aldehydes,
alcohols and quaternary
ammonium compounds,
such as Buraton 10 F
(Schülke & Mayr GmbH)
or Incidin Perfect
(Henkel Hygiene GmbH)
can be used to ensure
compatibility.
4)
Do not use any agents
containing alcohol.
5)
Disinfect the flow
sensor in 70% ethanol
solution for approx.
1 hour and leave
to dry in air for at least
30 minutes.