Alarm info
This function is used to list all active alarms / warnings in
order of priority.
● Press and hold the softkey »alarm info«.
Display (example):
Active alarms:
FIO2 <<
<<
!!!
Active warning:
ETCO2 <<
<<
!!
The display only remains visible as long as the softkey is
pressed.
CO2 alarms on /off
The alarm limits for etCO2, insp. CO2 and CO2 apnoea
monitoring can be switched off with the softkey
»CO2-al. on/off« in MAN/SPONT mode.
In this case, the symbol CO2 g appears alongside the
measured end-expiratory CO2 concentration.
The CO2 alarm limits are reactivated by pressing the
softkey again. The deactivated CO2 alarm limits are
reactivated automatically when changing to a different
ventilation mode.
The CO2 alarm limits can also be activated and
deactivated under "Alarm limits" in the configuration
screen in all ventilation modes, see page 72.
National and European standards require a minimum
monitoring with alarm function. These standards may not
be met if the alarm function of the etCO2 monitoring
parameter is deactivated.
Monitoring
CO2 alarms on/off
Alarm info
64
Pleth
IPPV
CO
2
40
0
PAW
0
20
600 10
1:2
10 0251.0050
ETCO2 !!
FIO2 !!!
FIO2 !!!
alarm
limits
auto set
limits
alarm
info
list
curves
config.
O
2
%
L/min
PEEP
mbar
Freq.
1/min
V
T
mL
T
I
:T
E
T
IP
:T
I
%
Pmax
mbar
Freshgas O
2
+ N
2
O
MAN/SPONT
CO
2
40
0
40 36
0.8 0.6
58 56
38
98 67
5.2 8
4.0030
VT
0.35
0
Volumeter
--s
Start volumeter:
Confirm !
--
05
10
0.5
1
Pleth
SpO
2
etCO
2
freq
MV
O
2
Hal.
N
2
O
Fi Fet
alarm
limits
CO
2-al.
on /off
alarm
Info
list
curves
config.
O
2
%
L/min
Freshgas O
2
+ N
2
O
fresh gas
_
CO
2
Ve
internal
external
"Only for internal use by Draeger. Not applicable as Instructions for Use!"