Preparing Julian for the self-test
Make sure that the breathing system is securely
connected.
1 Close the Y-piece = plug firmly on to the cone.
2 Ensure that the sample line is connected to the
Y-piece and to the water trap on the back of the
workstation.
Self-test
If all points on the checklist are OK:
Confirm = press the rotary knob.
The self-test is started.
The self-test is run automatically
and takes 3 to 4 minutes.
Julian carries out the following automatic tests and
actions:
Electronics
Testing and calibration of gas metering bench
Testing and calibration of O2 sensor
Testing of pressure sensor for airway pressure
measurement
Testing the flow sensor
Testing the battery
Activating the default settings for alarm limits,
monitoring parameters and the default settings of
the ventilator and fresh gas settings
Fresh gas mixer
Testing of gas supply O2, N2O, AIR
Testing of gas inlet valves
Calibration of pressures sensors in the compressed
air supply
Testing of O2 emergency metering
Testing of fresh gas failure monitoring (optional)
Ventilator and breathing system
Testing and calibration of PEEP/Pmax valves
Testing of safety valve
Testing of MAN/AUTO switch-over valve
Calibration of flow monitoring valve
Testing of time monitoring valve
Testing of bellows
31
Starting Up
Checking the workstation against the checklist
2
2
1