System compliance
Julian determines the current system compliance.
Depending on the breathing hoses used, the system
compliance is 5 to 6 mL/bar.
Leakage
Julian determines the current leakage of the breathing
system and breathing hoses.
The system tolerates leaks of up to 150 mL/min.
For leaks of more than 150 mL/min:
● Check the breathing system and repeat the leak test.
Display (example):
A tick (✓) indicates that the relevant test point has been
completed successfully.
The clock symbol uu
uu
indicates the test stage is currently
in progress.
Malfunctions detected in the self-test and any gas failures
are displayed on the screen. Some malfunctions can be
accepted by confirming = pressing the rotary knob, for
example no AIR supply (AIR FAILURE!!!).
Other malfunctions need to be corrected before starting
up the workstation, for example no O2 supply
(O2 FAILURE!!!).
Julian switches to standby after the self-test.
Display:
Starting Up
Checking the workstation against the checklist
32
Self-test
electronics
freshgas
mixer
ventilator
and
breathing system
system
compliance
leakage
✓
✓
✓
3.40 mL/mbar
20 mL/mbar
cancel
test
delete
trend
leakage
test
config.
Julian
SW-version 2.0
01. Nov. 98
Standby
4.00
L/min
30
Operation with , or
MAN
SPONT
IPPV
PCV
O
2
%
Freshgas O
2
+N
2
O