Alarms
Displaying and setting alarm limits
Alarms can be displayed and set from all three basic
screens (standard, data and trend screen) during
operation.
The default alarm limits for the ventilation mode are
activated automatically when changing from standby to a
ventilation mode.
These default alarm limits can be used unchanged or they
can be set individually for the patient.
The individual settings are deleted on changing from
ventilation mode to standby and the default alarm limits
are reactivated.
To call up alarm limits:
Press the softkey »alarm limits«.
Display (example):
The measured values and their alarm limits are shown on
the right-hand side.
The curves with the alarm limits represented by dashed
lines are shown on the left.
Example etCO2:
The upper alarm limit (42 mmHg) and the lower alarm
limit (30 mmHg) are shown alongside the measured value
(38 mmHg).
A deactivated alarm limit is indicated by two dashes (– –).
To set an alarm limit:
Place the cursor bar on the alarm limit by turning the
rotary knob and press to confirm.
The alarm limit is highlighted.
Set the new value by turning the rotary knob and
press to confirm.
The new alarm limit is now effective. The cursor bar
returns to the zz
zz
symbol.
Monitoring
Alarms
60
Pleth
IPPV
CO
2
40
0
600 10
1:2
10 0251.0050
98
67
38
6.0
29
0.8
--
95
3.0
--
20
1.0
0.0
40
8
120
50
--
Adjust limit value: Turn and confirm !
42
30
PAW
0
20
FiO
2
FiHal.
PAW
SpO
2
etCO
2
MV
alarm
limits
auto set
limits
alarm
info
list
curves
config.
O
2
%
L/min
PEEP
mbar
Freq.
1/min
V
T
mL
T
I
:T
E
T
IP
:T
I
%
Pmax
mbar
Freshgas O
2
+ N
2
O
38
42
30
etCO
2
"Only for internal use by Draeger. Not applicable as Instructions for Use!"