Intended Use
Julian Anaesthetic Workstation for patients with a body
weight of 5 kg and over with the use of IPPV ventilation.
For the use of:
Inhalation anaesthesia in rebreathing systems
Inhalation anaesthesia in semi-closed to virtually
closed systems with "low flow" and "minimal flow"
techniques (for minimal gas and anaesthetic agent
consumption).
Inhalation anaesthesia in non-rebreathing systems,
with separate fresh gas outlet for the connection of
e.g. the Bain system or Magill system
with a fresh gas flow of 0.5 to 12 L/min.
Ventilation modes:
Automatic ventilation (IPPV)
and
pressure-controlled ventilation (PCV).
Manual ventilation (MAN).
Spontaneous ventilation (SPONT).
The following measured values are displayed:
Peak pressure, mean pressure, plateau pressure and
PEEP
Expiratory minute ventilation
Tidal volume VT
Breathing rate
Patient compliance
Inspiratory and expiratory concentration of O2, N2O,
anaesthetic gas and CO2
optional:
Functional oxygen saturation (SpO2) and pulse rate.
The following parameters are displayed as curves:
Airway pressure
Expiratory flow
Inspiratory and expiratory concentration of O2, CO2
and anaesthetic gas
optional:
Plethysmogram
Trend curves and measured value lists are also available.
Monitoring
by means of adjustable alarm limits that are automatically
adapted to the ventilation mode.
The workstation may only be used under the
supervision of qualified medical personnel, so that
assistance can be provided immediately in the event
of any malfunctions.
Explosive anaesthetics, such as ether or cyclo-
propane, must not be used due to the risk of fire.
Mobile radio telephones must not be used within
10 metres of the workstation!
Mobile telephones may interfere with the operation of
electrical and electronic medical equipment.
Julian must not be used with nuclear spin
tomography (MRT, NMR, NMI)!
Operation of the apparatus may be impaired.
Intended Use
9
Intended Use
"Only for internal use by Draeger. Not applicable as Instructions for Use!"