Switch off the external fresh gas output:
● Press the softkey »Fresh gas internal/external« and
confirm by pressing the rotary knob.
Julian switches over to MAN/SPONT ventilation mode
or
● Press the »IPPV« or »PCV« key and confirm with
the rotary knob.
Julian then switches over directly to controlled
ventilation via the original rebreathing system.
When changing from the external non-rebreathing system
to the rebreathing system of the Julian:
● Reconnect the sample line to the Y-piece.
In the event of a power failure
Julian automatically switches over to the integrated
uninterruptible power supply (UPS).
If the battery is charged, operation will be maintained for
an average of 30 minutes.
The message
POWER FAIL!
appears on the screen
together with the remaining available battery capacity in
percent.
Example:
JJ
JJ
95 %
If the battery is almost completely discharged, the
message BATTERY LOW !!! appears on the screen.
In the event of a power failure if the battery is discharged,
Julian switches over automatically to MAN/SPONT
ventilation mode. The fresh gas mixer is deactivated.
Version without rotary knob for O2 emergency
metering:
Julian automatically delivers an O2 flow of 4 L/min.
This O2 flow bypasses the Vapor unit, and so the patient
no longer receives any anaesthetic agent.
Version with rotary knob for O2 emergency metering:
● Set rotary knob »Safety-O2« for O2 emergency
metering to the desired O2 flow.
Range 0 to 12 L/min.
This O2 flow flows through the Vapor unit.
● Check Vapor unit setting.
● Ventilate patient manually!
Operation
Using non-rebreathing systems
In the event of a power failure
52
Julian
D
Pleth
IPPV
POWER FAIL !
alarm
limits
95 %
auto set
limits
alarm
info
list
curves
CO
2
40
0
PAW
0
20
config.
O
2
SpO
2
J
N
2
O
Hal.
40 36
0.8 0.6
58 56
Fi
Fet
etCO
2
38
98 67
6.0 10
MV
freq
600
V
T
mL
10
1:2
10 0
PEEP
mbar
Freq.
1/min
T
I
:T
E
T
IP
:T
I
%
25
Pmax
mbar
Freshgas O
2
+N
2
O
1.00
O
2
%
L/min
50
_
CO
2
Ve
"Only for internal use by Draeger. Not applicable as Instructions for Use!"