Setting the measuring parameters
● Select the line »parameters« with the rotary knob and
press to confirm.
The »parameters« menu appears in the left-hand field.
Display (example):
Switch SpO2 measurement (optional) on/off:
● Select the line »SpO2 measurement« with the rotary
knob and press to confirm.
● Select »on« or »off« with the rotary knob and press to
confirm.
Select the sample rate for gas measurement:
● Select the line »sample rate« with the rotary knob and
press to confirm.
● Select »60« or »200 mL/min« with the rotary knob
and press to confirm.
For paediatric use:
● Select a sample rate of 200 mL/min in order to obtain
more accurate values on account of the high frequen-
cies and time constant of the measuring system.
● Recirculate the measured gas, see page 109.
If the measured gas is not recirculated:
● Increase the minute volume accordingly in order to
maintain the volume balance.
For adult use:
● Select a sample rate of 60 or 200 mL/min.
Recommendation: select 200 mL/min and recirculate
the measured gas, see page 109.
Set CO2 unit:
● Select the line »CO2 unit« with the rotary knob and
press to confirm. The last value selected is highlighted
by a grey background.
● Select the required unit of measure by turning the
rotary knob and press to confirm.
IPPV
0 1 2 3 4 5 6 7 8 9
1 2 3 4 5 6 7 8 9
adult child
600 10
1:2
10 0251.0050
O
2
%
L/min
PEEP
mbar
Freq.
1/min
V
T
mL
T
I
:T
E
T
IP
:T
I
%
Pmax
mbar
Freshgas O
2
+ N
2
O
PAW
40
20
0
settings
puls tone
alarm sound
Scale
parameter
record
activate monitor defaults
SpO
2
-measure.
sample rate
CO
2
unit
on off
60 200 mL/min
mmHg kPa vol%
Menu for configuration of measurement parameters.
anesth. agent
calibration
alarms
70
Monitoring
Configuring in operation
Setting the monitoring functions