End of Operation
To switch Julian to Standby:
Press the standby key E and confirm by pressing
the rotary knob.
The workstation is now on standby.
The fresh gas flow is switched off.
Manual ventilation is not possible!
The following default values are activated in standby
mode:
Fresh gas default values
Ventilation parameters
Alarm limits
To switch Julian off:
Press the power switch » « down completely.
The apparatus is switched off after 10 seconds.
Remove the gas supply probes from the wall sockets.
Leave Julian plugged into the power supply in order to
charge the uninterruptible power supply (UPS).
To switch off the anaesthetic gas scavenging system
AGS:
Pull back the black sleeve on the connector of the
suction hose to release the connector.
Julian is equipped with a switch-off time-lag. The
message "System will be closed down and switched off"
appears for 10 seconds after pressing the main power
switch. Julian can immediately be restarted by pressing
the main power switch again during this time.
The switch-off time-lag is not active during the self-test.