Setting the fresh gas default settings
● Select and confirm the line
»fresh gas default settings« with the rotary knob.
The softkeys for the fresh gas settings are displayed:
Display (example):
To modify the settings:
1 Press the keys »N2O« or »Air« to select the default
carrier gas and press the rotary knob to confirm.
2 Press the softkey »O2 %«.
Set the default O2 concentration with the rotary knob
and press to confirm.
3 Press the softkey »L/min«.
Set and confirm the default fresh gas flow with the
rotary knob.
To return to the »fresh gas default settings« menu:
● Press the rotary knob.