Connecting the power supply
Connecting auxiliary systems
1 connect to auxiliary sockets on the back of the
workstation.
The auxiliary sockets are not powered by the
uninterruptible power supply UPS in the event of a
power failure!
Do not connect HF surgical devices to the auxiliary
sockets!
Connecting equipment to the auxiliary sockets may cause
the patient leakage current to rise above the permitted
values in the event of failure of a protective earth
conductor.
The risk of electric shock cannot be excluded in such
cases.
Additional power adapter sockets must not be connected
to the auxiliary sockets.
Equipotential bonding
e.g. for intracardial or intracranial operations.
2 Connect one end of the earth cable to one of the
connecting pins on the back of the workstation.
Connect the other end of the earth cable to the
specified equipotential bonding point, e.g. on the
operating table or ceiling lamp.
Connecting the power supply
The mains voltage must be the same as specified on the
nameplate on the back of the workstation.
Voltage range: 90 to 265 V
Plug the mains plug into the wall socket.
24
Preparation
Connecting the power supply
1
2