Dismantling the suction system
1 Remove the suction hose and vacuum hose.
2 Grip the silicone sleeve of the bottle cap and pull it
off.
3 Remove the secretion collecting bottle and flushing
bottle from the bracket and empty them.
Note the hygiene regulations of the hospital.
4 Remove the silicone sleeve from the lid of the
secretion collecting bottle.
5 Remove the overflow protection float from the rising
pipe.
Store the parts together for disinfection and cleaning
in a washing machine.
6 Replace the bacterial filter in the back panel of the
Julian – if soaked through or if contamination is
suspected.
The filter must be replaced at the latest every two
weeks.
● The bacterial filter must be disposed of as infectious
special waste. It can be incinerated with little pollution
at temperatures of more than 800 °C.
2
33
1
1
98
Care
Dismantling components
4
5
6