Dismantling the suction system
1 Remove the suction hose and vacuum hose.
2 Grip the silicone sleeve of the bottle cap and pull it
off.
3 Remove the secretion collecting bottle and flushing
bottle from the bracket and empty them.
Note the hygiene regulations of the hospital.
4 Remove the silicone sleeve from the lid of the
secretion collecting bottle.
5 Remove the overflow protection float from the rising
pipe.
Store the parts together for disinfection and cleaning
in a washing machine.
6 Replace the bacterial filter in the back panel of the
Julian – if soaked through or if contamination is
suspected.
The filter must be replaced at the latest every two
weeks.
The bacterial filter must be disposed of as infectious
special waste. It can be incinerated with little pollution
at temperatures of more than 800 °C.
2
33
1
1
98
Care
Dismantling components
4
5
6