Care
Dismantling components
Dismantling sample line
Unscrew the sample line from the Y-piece and water
trap on the back of the unit.
The sample line is not reusable and can be disposed
of with ordinary domestic waste.
Removing water trap container
Pull the container of the water trap down and off, and
empty it.
Note the hygiene regulations of the hospital!
Prepare the container of the water trap for disinfection
and cleaning in a washing machine.
When using the "Waterlock" water trap
See separate Instructions for Use.
Removing breathing hoses
Remove the breathing hoses and breathing bag.
Dismantle Y-piece and connector.
Prepare the breathing hoses, breathing bag and
Y-piece for disinfection and cleaning in a washing
machine.