Setting the monitoring functions
● Select the »settings« column by turning the rotary
knob and press to confirm.
Display (example):
Setting the pulse tone
0 = OFF
9 = Maximum volume
● Select the line »pulse tone« by turning the rotary knob
and press to confirm. The cursor automatically jumps
to the last value selected (in the example: setting 5).
● Set the value with the rotary knob and press to
confirm.
Setting the alarm sound
1 = Minimum volume
9 = Maximum volume
● Select the line »alarm sound« by turning the rotary
knob and press to confirm. The cursor automatically
jumps to the last value selected (in the example:
setting 4).
● Set the value with the rotary knob and press to
confirm.
Set a sufficiently loud alarm sound!
The audible »NO O2« and »NO FRESH GAS« alarms are
always sounded at maximum volume.
Julian complies with national regulations in certain
countries that stipulate a minimum volume of 45 dB (A).
Settings 1 to 4 are programmed at 45 dB (A) for these
countries.
Selecting the scale
To adapt the scale for the volumeter function and trend
curve.
● Select the line »scale« with the rotary knob and press
to confirm.
● Select »adult« or »child« with the rotary knob and
press to confirm.
69
Monitoring
Configuring in operation
Setting the monitoring functions
O
2
-Sensor
21 Vol.%
gas sensor
IPPV
anesth. agent
calibration
pulse tone
alarm sound
scale
parameters
record
activate monitor defaults
0 1 2 3 4 5 6 7 8 9
1 2 3 4 5 6 7 8 9
adult. child
alarms
600 10
1:2
10 0251.0050
✓
✓
O
2
%
L/min
PEEP
mbar
Freq.
1/min
V
T
mL
T
I
:T
E
T
IP
:T
I
%
Pmax
mbar
Freshgas O
2
+ N
2
O
selection
auto man
Standard
CO
2
on off
SpO
2
on off
HLM on off
Isoflurane
PAW
40
20
0
The anesthetic
agent is automati-
cally selected by
gas identification
settings
"Only for internal use by Draeger. Not applicable as Instructions for Use!"