Operation
Using non-rebreathing systems
Display (example):
The airway pressure (PAW), minute volume (MV) and
frequency are not measured.
Set the fresh gas flow. The fresh gas supply must be
at least twice the minute volume in order to prevent
rebreathing.
Certain alarms are automatically switched off in order to
avoid artefacts.
The inactive alarm limits are highlighted by the grey
background in the following table.
FRESHGAS EXTERNAL
CO
2
40
0
30 28
0.8 0.8
58 57
38
98 67
--
--
6.0030
Pleth
O
2
SpO
2
N
2
O
Hal.
Fi
Fet
etCO
2
MV
freq
CO
2 al.
on/off
alarm
limit
alarm
info
list
curves
config.
Freshgas O
2
+ N
2
O
O
2
%
L/min
fresh gas
_
CO
2
Ve
internal
external
Alarm limits In external fresh
gas mode
Default setting
on delivery
SpO2 ON
ON
– –
92
Pulse
H
ON
ON
120
50
etCO2 ON
ON
50 (mmHg)
– –
FiCO2 ON Fixed upper alarm
limit 5 mmHg
MV OFF
OFF
FiO2 OFF
ON
20
Fi Hal. ON
OFF
1.5
– –
Fi Iso. ON
OFF
2.3
– –
Fi Enf. ON
OFF
3.4
– –
Fi Des. ON
OFF
12.0
– –
Fi Sev. ON
AUS
3.4
– –
PAW ON
OFF
40
– –
Apnoea pressure OFF
Apnoea flow OFF
Apnoea CO2 ON
_
_
_
>
_
_
_
_
_
_
_
_
"Only for internal use by Draeger. Not applicable as Instructions for Use!"